SDS5000X Series Digital Oscilloscope User Manual
15.5.8 Dropout Trigger
Dropout trigger includes two types: Edge and state.
Edge
Trigger when the time interval (â–³T) from when the rising edge (or falling edge)
of the input signal passes through the trigger level to when the neighboring
rising edge (or falling edge) passes through the trigger level is greater than the
set time, as shown in the figure below:
State
Trigger when the time interval (â–³T) from when the rising edge (or falling edge)
of the input signal passes through the trigger level to when the neighboring
falling edge (or rising edge) passes through the trigger level is greater than the
set time, as shown in the figure below:
Trigger source, slope (rising, falling), dropout type and time value can be set in
the trigger dialog box. Holdoff, coupling and noise reject can be set in dropout
trigger, see the sections " Holdoff", " Trigger Coupling" and " Noise Reject" for
details.