SDS5000X Series Digital Oscilloscope User Manual
Label
Users can set the label text for channels. Open the label setting dialog box to
select the source, customize the label text, and set the display. The source can
be C1~C4, Math and RefA~RefD. The length of the label is limited to 20
characters. The characters beyond this length will not be displayed. When the
“Display” option is set toon, the label will be displayed on the right side of the
channel offset indicator.
The labels are hidden
The labels are displayed
Applied to
With this setting, some common functions such as Cursor, Measure, FFT,
Search, DVM, Histogram, Mask Test and AWG can be quickly applied to the
selected channel. Once a function is specified, it will switch directly to the
function menu and automatically set that channel as the source.
Impedance
1 : When a passive probe with high impedance is connected, the
impedance must be set to 1 , otherwise the signal will not be detected.
50 Ω: Suitable for high frequency signals transmitted through 50 Ω coaxial
cables and can minimize the amplitude distortion caused by impedance
mismatching.