SDS5000X Series Digital Oscilloscope User Manual
each zone as intersect or not intersect as an additional condition to further
isolate the interested event quickly. “Intersect” only includes events that occur
within the zone. “Not-intersectevents include all events that occur outside of
the zone.
Touch Trigger>Zone to recall zone dialog box:
A. Turn on or off zone trigger
B. Select the source: C1~C4
C. Turn on or off zone1
D. Set the property of zone1: Intersect or
Not intersect
E. Set the coordinate of zone1. The range is
within the waveform area
F. Turn on or off zone2
G. Set the property of zone2: Intersect or
Not intersect
H. Set the coordinate of zone2. The range is
within the waveform area
The zones can be created and moved by gestures or by setting Zone> Zone
Setting in the dialog box. The color of the zone’s outline is consistent with the
color of the specified source (Channel 1 = Yellow, etc...).