SDS5000X Series Digital Oscilloscope User Manual
90 / 2 3 6 W W W. S I G L E N T. C O M
15.5.4 Pulse Trigger
Trigger on a positive or negative pulse with a specified width. Trigger source,
polarity (positive, negative), limit range and time value can be set in the trigger
dialog box.
Less than a time value
-- Trigger when the positive or negative pulse
time of the input signal is lower than the specified time value. Below is an
example of trigger condition is set to positive pulse width < 100ns
Great than a time value
-- Trigger when the positive or negative pulse
time of the input signal is greater than the specified time value. Below is an
example of trigger condition is set to positive pulse width > 100ns
Within a range of time value
[--,--]-- Trigger when the positive or negative
pulse time of the input signal is greater than the specified lower limit of time and
lower than the specified upper limit of the time value. Below is an example of
trigger condition is set to 100 nspositive pulse width300 ns.