Manual - Page 191

For SDS5052X.

Loading ...
Loading ...
Loading ...
SDS5000X Series Digital Oscilloscope User Manual
191 / 236 W W W. S I G L E N T. C O M
anomalies in the waveform can be found in short time without complex trigger
settings to improve test efficiency. Below is an example to display glitches in a
data sequence with infinite persistence.
When the display dialog box and persist are off, double pressing the
Display/Persist button on the front panel can quickly turn on persist.
Touch Persist in the display dialog box to set the persistence time.
Off: Turn off persist.
Variable persist time (1s, 5s, 10s, 30s): Choose different persist
times. The oscilloscope updates with the new acquired waveform
display. Acquired waveforms will be cleared after the corresponding
time has expired.
Infinite: Select Infinite, previous acquisitions will never be erased
until a Clear Display or Clear Sweeps operation is performed.
Loading ...
Loading ...
Loading ...