SDS5000X Series Digital Oscilloscope User Manual
199 / 236 W W W. S I G L E N T. C O M
OVP
When OVP is enabled, the output will be turned off automatically once the
protection condition is met. The protection condition is when the absolute value
on the output port is higher than 4 V ± 0.5 V). At the same time, a warning
message is displayed.
Sync Output
When the Sync Output is turned on, the Aux In/Out port of AWG outputs a
CMOS signal with the same frequency as the basic waveform (except Noise
and DC) and arbitrary waveform. The maximum supported frequency is 10 MHz.
Zero Adjust
The output zero error of the AWG can be calibrated by using auto mode or
manual mode. The object of the Zero Adjust is: when AWG outputs 0 V DC
signal, the measured mean value of the corresponding channel is within ±1
mV.
Auto Steps: In this mode the adjustment can only be performed with CH1
of the oscilloscope. Connect the output of AWG to CH1 of the oscilloscope,
press Auto, and the AWG will starts to automatic adjust. The scope will
displayZero adjust completed! when the adjustment is done.
Manual Steps: The AWG can be manually calibrated through any channel
on the oscilloscope. Take CH2 as an example:
1. Connect the output of AWG to CH2, open CH2, set it to DC coupling,
turn on bandwidth limit, and set the probe attenuation to 1X.