SDS5000X Series Digital Oscilloscope User Manual
19 / 2 3 6 W W W. S I G L E N T. C O M
6 Probe
The SDS5000X series oscilloscope package includes passive probes as
standard accessories. Please visit the website at
www.siglent.com for
technical data and ordering information.
Probe Compensation
When a probe is used for the first time, you should compensate it to match the
input channel of the oscilloscope. Non-compensated or poorly compensated
probe may increase measurement inaccuracy or error. The probe
compensation procedures are as follows:
1. Connect the coaxial cable interface (BNC connector) of passive probe
to any channel of the oscilloscope.
2. Connect the probe to the “Compensation Signal Output Terminal
(Cal) on the front of the oscilloscope. Connect the ground alligator clip
of the probe to the “Ground Terminal” under the compensation signal
output terminal.
3. Press the Auto Setup button.
4. Check the waveform displayed and compare it with the following.