Period: Time between the middle threshold points of two
consecutive like-polarity edges.
Freq: Reciprocal of period
+Width: Time difference between the 50% threshold of a rising
edge to the 50% threshold of the next falling edge of the pulse
-Width: Time difference between the 50% threshold of a falling
edge to the 50% threshold of the next rising edge of the pulse
Rise: Duration of rising edge from 10-90%
Fall: Duration of falling edge from 90-10%
BWidth: Time from the first rising edge to the last falling edge or
the first falling edge to the last rising edge at the 50% crossing
+Duty: Positive Duty Cycle. Ratio of positive width to period
-Duty: Negative Duty Cycle. Ratio of negative width to period
Delay: Time from the trigger to the first transition at the 50%
crossing
T@M: Time from the trigger to each rising edge at the 50%
crossing
18.3.3 Delay Measurement
Delay measurement measures the time difference between two channels. It
includes 10 delay parameters: