15.5 Trigger Type
15.5.1 Overview
The trigger modes of the SDS5000X are digital designs. Compared with analog
trigger circuits, digital triggers can not only greatly optimize trigger precision,
trigger jitter, but also support multiple trigger types and complex trigger
conditions.
Edge -- Trigger on rising edge, falling edge or both
Slope -- Trigger when an edge crosses two thresholds which lie inside or
outside a selected time range
Pulse -- Trigger at the end of a pulse when the pulse width lies inside or
outside a selected time range
Video -- TV trigger on falling edge
Window -- Trigger when the signal leaves the widow region.
Interval -- Trigger on second edge when the time between the edges is
inside or outside a selected time range
Dropout -- Trigger when signal disappears for longer than Dropout value.
Runt -- Trigger when a pulse crosses the 1st threshold but not the 2nd
before re-crossing the 1
st
threshold again
Pattern -- Trigger when pattern condition transitions from false to true. All
inputs set to DC coupling
Serial -- Trigger on specified condition in a serial bus. See the chapter
"Noise Reject = On
Serial Trigger and Decode " for details.
Qualified -- Trigger with edge trigger setting only after the qualifying
condition is satisfied