Chapter 2 To Set the Vertical System RIGOL
MSO2000A/DS2000A Users Guide 2-5
I nput I mpedance
This oscilloscope provides two input impedance modes (1(default) and 50 Ω) to
reduce the circuit load caused by the interaction of the oscilloscope and the circuit to
be tested.
1MΩ: the input impedance of the oscilloscope is rather high and the current
flowing into the oscilloscope from the circuit under test can be ignored.
50Ω: match the oscilloscope with a device with 50 Ω output impedance.
Press CH1 I nput to set the input impedance of the oscilloscope. When “50Ω” is
selected, Ω is displayed in the channel status label at the bottom of the screen.
Note: The setting of the input impedance affects the settable range of the vertical
scale and vertical offset of the corresponding channel. For the details, please refer to
Specifications”.
Waveform I nvert
Press CH1 I nvert to turn waveform invert on or off. When waveform invert is
turned off, the waveform display is normal; when waveform invert is turned on, the
waveform voltage values are inverted (as shown in the figures below).
(a) Invert is Off
(b) I nvert is On
Figure 2-1 Waveform I nvert