RIGOL Chapter 1 Quick Start
1-8 MSO2000A/DS2000A Users Guide
Probe Compensation
When the probes are used for the f irst time or when the probe compensation signal
does not match that shown in Figure 8-1, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
compensated probes may cause measurement error. The probe compensation
procedures are as follows.
1. Perform steps 1, 2, 3 and 4 of Function I nspection”.
2. Check the displayed waveforms and compare them with the following figures.
Figure 1-9 Probe Compensation
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment
hole on the probe until the displayed waveform is as the Perfectly
compensated in the figure above.
Over compensated Perfectly compensated Under compensated