RIGOL Chapter 5 To Trigger the Oscilloscope
5-16 MSO2000A/DS2000A Users Guide
upper limit and Low Level denotes the trigger level lower limit. The trigger level lines
and trigger marks disappear after you stop rotating the knob for about 2 s.
The adjustment mode of the trigger level differs when different vertical window is
selected.
: only adjust the upper limit of the trigger level. During the adjustment,
the Up Level” changes accordingly and Low Level remains unchanged.
: only adjust the lower limit of the trigger level. During the adjustment,
the Low Level” changes accordingly and the Up Level” remains unchanged.
: adjust the upper and lower limits of the trigger level at the same time.
During the adjustment, the Up Level” and Low Level change accordingly.
Trigger Mode:
Press Sweep and select Auto”, Normal or Single”. For the details, please refer to
Trigger Mode. The corresponding status light of the current trigger mode turns
on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.