RIGOL Chapter 18 Specifications
MSO2000A/DS2000A User’s Guide
18-2
I nput
Number of
Channels
MSO2XX2A/ 2XX2A-S: 2 analog channels+ 16 digital channels
DS2XX2A/ 2XX2A-S: 2 analog channels
Input Coupling DC, AC or GND
Input Impedance
Analog channe: (1 MΩ±1% ) | | (16 pF± 3 pF) or 50 Ω± 1.5%
Digital channel: (101 kΩ±1%)| | (9 pF± 1 pF)
Probe
Attenuation
Coefficient
Analog channe: 0.01X to 1000X, in 1-2-5 step
Maximum I nput
Voltage (1 MΩ)
Maximum Input Voltage of the Analog Channel
CAT I 300 Vrms, CAT I I 100 Vrms,
Transient Overvoltage 1000 Vpk
Digital channel: CAT I 40Vrms, Transient Overvoltage 800 Vpk
Horizontal
Time Base Scale MSO/ DS2302A/2302A-S: 1.000 ns/ div to 1.000 ks/div
MSO/DS2202A/2202A-S: 2.000 ns/div to 1.000 ks/div
MSO/DS2102A/2102A-S/ 2072A/ 2072A-S:
5.000 ns/div to 1.000 ks/div
Deviation
between
Channels
1 ns (typical), 2 ns (max)
Max Record
Length
14 Mpts
Time Base
Accuracy
[1]
≤ ± 25 ppm
Clock Drift ≤ ± 5 ppm/ year
Max Delay Range Negative delay: ≥1 screen width
Positive delay: 1 s to 100 ks
Time Base Mode Y-T, X-Y, Ro l l
Number of XYs 1
Waveform
Capture Rate
[2]
50,000 wfms/ s (dots display)
Loading ...