Chapter 6 MATH and Measurements RIGOL
MSO2000A/DS2000A Users Guide 6-3
Press Label to define a label for the operation results. For the details, please
refer to
MATH Label”.
HORI ZONTAL
POSI TI ON and HORI ZONTAL SCALE can also be
used to adjust the horizontal position and scale of the operation results.
Multiplication
Multiply the waveform voltage values of signal source A and B point by point and
display the results. Press MATH Operate to select A x B:
Press Source A and Source B respectively to select the desired channels (CH1
or CH2).
Press
and use to adjust the vertical position of the operation results.
Press
and use to adjust the vertical scale of the operation results.
Press I nvert to turn the inverted display of the operation results on or off.
Press Label to define a label for the operation results. For the details, please
refer to
MATH Label”.
HORI ZONTAL
POSI TI ON and HORI ZONTAL SCALE can also be
used to adjust the horizontal position and scale of the operation results.
Division
Divide the waveform voltage values of signal source A by that of source B point by
point and display the results. It can be used to analyze the multiple relationships of
waveforms in the two channels.
Note: When the waveform voltage value of signal source B is 0, the result of the
division is treated as 0.
Press MATH Operate to select A ÷ B” :
Press Source A and Source B respectively to select the desired channels (CH1
or CH2).
Press
and use to adjust the vertical position of the operation results.
Press
and use to adjust the vertical scale of the operation results.
Press I nvert to turn the inverted display of the operation results on or off.
Press Label to define a label for the operation results. For the details, please
refer to
MATH Label”.