RIGOL Chapter 18 Specifications
MSO2000A/DS2000A User’s Guide
18-12
General Specifications
Probe Compensation Output
Output Voltage
[1]
About 3 V, peak-peak
Frequency
[1]
1 kHz
Power
Power Voltage 100 V to 240 V, 45 Hz to 440 Hz
Power Maximum 50 W
Fuse 2 A, T Degree, 250 V
Environment
Tem p er at u r e Range
Operating: 0℃ to + 50℃
Non-operating: -40℃ to + 70℃
Cooling Method Fan cooling
Humidity Range
0℃ to + 30℃: ≤95% Relative Humidity
+ 30℃ to + 40℃: ≤75% Relative Humidity
+ 40℃ to + 50℃: ≤45% Relative Humidity
Altitude Operating: under 3,000 meters
Non-operating: under 15,000 meters
Physical Characteristics
Size
[3]
Width × Height × Depth = 361.6 mm × 179.6 mm ×
130.8 mm
Weight
[4]
Package Excluded 3.9 kg ± 0.5 kg
Package I ncluded 4.5 kg ± 0.5 kg
Calibration I nterval
The recommended calibration interval is one year.
Regulatory I nformation
Electromagnetic
Compatibility
2004/108/EC
Execution standard EN 61326-1:2006 EN 61326-2-1:2006
Safety UL 61010-1:2004; CAN/ CSA-C22.2 NO. 61010-1-2004;
EN 61010-1:2001; IEC 61010-1:2001
Note
[1]
: Typical.
Note
[2]
: Maximum value. 20ns, single-channel mode, dots display, auto memory depth.
Note
[3]
: Supporting legs and handle folded, knob height included.
Note
[4]
: Standard configuration.