RIGOL Chapter 6 MATH and Measurements
6-30 MSO2000A/DS2000A Users Guide
modes available are Manual, Tr ack , Auto and X-Y. When OFF is selected, the
cursor measurement function is turned off. When the Time Base Mode is X-Y,
only X-Y cursor mode is available.
Manual Mode
In this mode, you can adjust the cursors manually to measure the X value (or Y
value), X increment (or Y increment) between cursors and the reciprocal of X
increment on the waveform of the selected source (CH1, CH2 or MATH) . When the
measurement source is set to LA, the logic level value of the digital channel currently
turned on will be displayed in binary and hexadecimal forms (high level is 1 and low
level is 0).
Press Cursor Mode Manual to turn the manual cursor measurement
function on. The measurement results are displayed at the upper left corner of the
screen. You can also press Cursor or Mode continuously to switch to Manual”.
When the measurement parameters are modif ied, the measurement results will
change accordingly.
When the measurement source is CH1, CH2 or MATH, the measurement results are
as shown in the f igure below.
A-> X: the X value at cursor A. X value takes the trigger position as reference. Its
unit is s” or Hz (when measuring FFT waveform).
B-> X: the X value at cursor B. X value takes the trigger position as reference. I ts
unit is s” or Hz (when measuring FFT waveform).
A-> Y: the Y value at cursor A. Y value takes the channel GND as reference. I ts
unit is the same as that of the current signal source.
B-> Y: the Y value at cursor B. Y value takes the channel GND as reference. I ts
unit is the same as that of the current signal source.