Chapter 13 Signal Source RIGOL
MSO2000A/DS2000A Users Guide 13-5
To Output Ramp
Press Setup1 to open the waveform setting menu. Press Wave and select Ramp.
At this point, you can set the ramp parameters. For the detailed setting method, refer
to the introduction in
To Output Sine”. This section only introduces Symmetry”.
Symmetry
It is defined as the percentage that the rising period of the ramp takes up in the
whole period.
T
t
Symmetry= t/ T* 100%
Set the symmetry
Press Symmetry and use or the navigation knob to set the symmetry of the
ramp waveform. The range is from 0% to 100% . The default is 10.0% .