Pressurements PressCal
Users Manual
5-2
Note
If new values are manually entered they are only available for the current
session. When the session is restarted the values reset to default.
For information on how to adjust the default settings, refer to Setting the Default Values
for details.
5. Check:
Correct PCU shown within the “Piston S/No.” field. If the required PCU is not
Shown, use the pull-down list to view all PCU options for the current DWT, then
select. Alternatively, refer to Managing PCU Details.
Correct Weight Sets shown within the “Weight Set S/No.” field. Clicking on a
new option in the list causes the “Weight Selection Details” to update.
6. Adjust the following settings:
Note
If new values are manually entered they are only available for the current
session. When the session is restarted the values reset to default.
Tilt
Enter the angle of the DWT by either inputting the value
directly or by using the arrows (
v
) for fine adjustment.
Note
Where possible the DWT should be level, as
stated within the DWT’s operating instructions.
Pressure Head
Adjust the height of the pressure head or accept the
default. Refer to Setting the Default Values.
Piston / Cylinder
Temp
Due to the thermal characteristics of the PCU it may be
necessary to adjust the value of the piston/cylinder
temperature within PressCal. Please refer to the DWT
operating manual (or supplier’s information).
Resolution
Adjust the number of decimal places that you want the
Target Pressure and Computed Pressure to be
calculated to.
Pressure Units
Using the pull-down list select the preferred units,
e.g. bar,mbar, MPa, kPa, psi, kg/cm
2
etc.
Calculation
There are two methods of calculation:
Note
Both methods take into account all environmental and working settings.
Select a Target Pressure and allow PressCal to automatically select the
combination of weights from the Weight Set that will achieve the closest match to
the target pressure.
Select the preferred weight(s) from the “Weight Selection Details” and allow
PressCal to calculate the resultant pressure.