Setting the Default Values
The default values have been defined by SI Pressure Instruments. However, depending on
your environment and your testing specifications, these may need to be modified.
Click on Setup > Default Settings within the “Menu Bar” to open the “Default
Values” window as shown in Figure 2-5.
gmp06.bmp
Figure 2-5. Default Values Window
1. Enter the following:
Note
Either use the arrows (
v
) to finely adjust the default values, or
alternatively enter the values in directly.
Head
Subtract the Unit Under Test (UUT) height from the
reference standard height (i.e. if the UUT is above the
reference level then the Head value becomes -ve).
Default is zero.
Ambient Pressure
Enter the ambient pressure of the calibration test
environment. This can be obtained by using a suitable
manometer (such as SI Pressure’s PC6).
Ambient
Temperature
Enter the ambient temperature of the calibration test
environment. The default calibration temperature is 20 ºC.
Humidity
Enter the current humidity. This can be obtained by using
a suitable hygrometer.
2. Click on Save to store the default value changes or click on Exit to close the
window without saving the changes.