SDS1000X-E&SDS1000X-U User Manual
52
WWW.SIGLENT.COM
Figure 24 Pulse With 0.1% Duty, Peak Detect Mode
Average
In this mode, the oscilloscope averages the waveforms from multiple samples to
reduce the random noise of the input signal and improve the vertical resolution.
The greater the number of averages is, the lower the noise will be and the higher
the vertical resolution will be but the slower the response of the displayed
waveform to the waveform changes will be.
The available range of averages is from 4 to 1024 and the default is 16. When
Average mode is selected, press
Averages and turn the Universal knob or press the
softkey continually to set the desired average time.