SDS1000X-E&SDS1000X-U User Manual
129
WWW.SIGLENT.COM
window function according to different waveforms and their
characteristics. Please read the table below carefully to make an
appropriate option according to the input signal.
Window
Applications and Characteristics
Rectangle
These are normally used when the signal is transient
(completely contained in the time-domain window) or known
to have a fundamental frequency component that is an integer
multiple of the fundamental frequency of the window. Signals
other than these types will show varying amounts of spectral
leakage and scallop loss, which can be corrected by selecting
another type of window.
Hanning
These reduce leakage and improve amplitude accuracy.
However, frequency resolution is also reduced.
Hamming
These reduce leakage and improve amplitude accuracy.
However, frequency resolution is also reduced.
Flat Top
This window provides excellent amplitude accuracy with
moderate reduction of leakage, but with reduced frequency
resolution.
Blackman
It reduces the leakage to a minimum, but with reduced
frequency resolution.
Press the Auto set softkey to automatically set the appropriate parameters
for the FFT measurement.
Press the Display softkey to select Split Screen, Full Screen or Exclusive
display mode.
Split Screen: Time domain waveform and frequency domain waveform are
displayed separately. The time domain waveform is on the upper half
screen, while the frequency domain waveform is located within the lower
half of the display. In Split mode, if Zoom is enabled, the zoom waveform
and the frequency domain waveform are displayed on the lower half
screen together.
Full Screen: Time-domain waveform and frequency-domain waveform are
displayed together.
Exclusive: Only the frequency-domain waveform is displayed.
Press the Mode softkey to select Normal, Max-Hold or Average. When you
select Average, it is necessary to set the average times.
5. Press the Vertical softkey to enter VERTICAL menu.
Figure 75 VERTICAL Menu
Press the Scale softkey, and then turn the Universal Knob to select the
desired vertical FFT scale
Press the Ref Level softkey, and then turn the Universal Knob to select the