SDS1000X-E&SDS1000X-U User Manual
180
WWW.SIGLENT.COM
temperature is up to or more than 5 ℃. Make sure that the oscilloscope has been
warmed up or operated for more than 30 minutes before the self-calibration.
Do the following steps to do self-calibration:
1. Disconnect all the input channels.
2. Press the Utility button on the front panelïĵŒand then press the Do Self Cal
softkey, and the oscilloscope will pop-out the message box shown as below:
Figure 113 Do Self Cal
3. Press the Single button on the front panel to perform the self-calibration
program. During the calibration, most of the keys are disabled.
4. When the self-calibration program is finished, it will pop-out the message
“
Press Run/Stop key to exit”. Press the Run/Stop button on the front panel to
exit the calibration interface.