SDS1000X-E&SDS1000X-U User Manual
220
WWW.SIGLENT.COM
Reference Position
The reference position setting determines the physical point that the oscilloscope
uses during vertical and horizontal scale changes. In some situations, it is more
convenient to use a fixed position on the display.
Press the Utility button on the front panel and then press the Reference Pos. softkey
to enter the Reference POS menu.
Press the Vertical softkey to select Fixed Offset or Fixed Position.
Fixed Position: when the vertical gain is changed, the oscilloscope will keep the
vertical offset level indicator stationary.
Fixed offset: when the vertical gain is changed, the oscilloscope will have the
vertical offset level indicator move with the actual voltage level.
Press the Horizontal softkey to select Fixed Delay or Fixed Position.
Fixed Position: when the timebase is changed, the oscilloscope will keep the
horizontal offset indicator stationary.
Fixed Delay: when the timebase is changed, the oscilloscope will have the
horizontal offset indicator move with the trigger point.
Power On Line
Press the Utility button on the front panel; then; press Power On Line softkey to
select Enable or Disable.
Enable: when the oscilloscope is energized, the oscilloscope will start
automatically by itself.
Disable: when the oscilloscope is energized; then; press the power key at the
lower-left corner of the front panel to start the oscilloscope.