SPI Serial Decode
After completing the setup of SPI signal and trigger, we will decode SPI signals.
Operation steps as follows.
1. Press Decode Decode. Select one of the options from the Decode1 and
Decode2.
2. Press the Configure to set the bitstream format and bits (4-32 bits)
3. Press the Display and select On to display the result of decoding.
4. Press the Configure to set bit order and data length.
5. Press the List to enter the LIST function menu.
Press the Display and choose the same options as the first step.
Press the Scroll and turn the Universal Knob to view all frames.
Press the Lines and set the number of lines by Universal Knob. The range
of the lines is 1 to 7.
Users can export the result list of the package to the external USB storage
device in csv format (only when the USB flash disk is detected). Saving the
data (*.csv) is similar to the operation of setup files, see the section
"External save and recall" for details.
6. Press the Format to change the character encoding format of the decodings
result.
7. Press the Copy Setting to enter the COPY function menu to synchronize the
corresponding bus configuration and trigger configuration.
8. Press the Tips Info to turn on or off the decoding limit prompt. When the
decoding frame number reaches the maximum, it will pop up "Decoding to
maximum frame number limitation!"
Interpreting SPI Decode
The frames of decoding result:
The data values are displayed in frames and are shown in white. Support 4~96
bit data display.
MISO the decoding result of “Master-In, Slave-Out” line.
MOSI —the decoding result of “Master-Out, Slave-In” line.
Indicates there is not enough space on the display to show the complete