44 UTD1000L User Manual
Figure 5-20 Measuring AC Current Greater than 400mA
5.5 Data Hold
You can freeze the displayed readings any time as you want.
1. Press RUN /STOP to freeze measuring values, and HOLD icon blinks on the display
2. Repress RUN /STOP to restore the measurement
Figure 5-21 Freezing Measurement Values
5.6 Making Relative Measurement
Relative measurement displays a current measuring result relative to the specific reference
value.
Refer to capacitance measurement in relative mode. First of all you need to obtain a reference
value:
1. Press R button to set measurement type to Resistance.
2. Press F1 button to select Capacitance option
3. Connect test leads to input terminals (Black to COM and Red to V/Ω)
4. Wait until the reading becomes stable, then press F2 button to access the relative mode,
icon shows on the upper display and the reference value appears just below icon.
5. Connect test leads to the capacitor under test, then take capacitance reading on the display.