40 UTD1000L User Manual
2. Press F3 to select mA range, and the measurement unit will display mA accordingly.
3. Plug mA current module (UT-M10) into input terminalsCOM and V/Ω, then turn the switch
on the module to 400mA range.
4. Insert Black and Red test leads correspondingly into Black and Red terminals on the module.
5. Then connect test leads to tested points, than take the DC current reading.
Figure 5-15 Measuring DC Current within 40mA-400mA
To measure DC current greater than 400mA, do the following:
1. Press I button to set measurement type to DC Current.
2. Press F3 to select A range, and the measurement unit will display A accordingly.
3. Plug 4A current module (UT-M04) into input terminals COM and V/Ω, then
insert test leads properly into the module.
4. Connect test leads to tested points, then take DC current reading.
Warning: If test leads are directly connected to V/Ω and COM terminals without use of 4A
current module (UT-M04) when measuring the current greater than 400mA, the fuse in the
product will be burnt out. Please have the fuse replaced by qualified personnel if the incident
happens.