UNI-T UTD1025DL Multimeter/Oscilloscope

UTD1000L Series user manual - Page 20

For UTD1025DL.

PDF File Manual, 91 pages, Read Online | Download pdf file

UTD1025DL photo
Loading ...
Loading ...
Loading ...
18 UTD1000L User Manual
Figure 4-4 Adjusting Probe Factor
3. Conenct the probe to the output terminal for probe compensation on the Oscilloscope,
press AUTObutton, after several seconds, squarewave displays (1kHz, 3Vpp, Peak-to-Peak
Value), See Figure 4-5.
Figure 4-5 Squarewave Signal from the Oscilloscope Itself
4.8 Compensating the Probe
You need to compensate the probe whenever connecting it to channels for the first time so as
to match the probe to the input channel. Uncompensated probe may cause deviation or error
to the measurement. To compensate the probe, do the following:
1. First of all, set probe factor on the Oscilloscope to 10×, turn the switch to 10× on the probe,
then connect the probe connector to the oscilloscope input channel and the probe tip to
compensation output terminal;
2. Then execute AUTO
3. Observe the displayed waveform
Over Compensated Correctly Compensated Under Compensated
Loading ...
Loading ...
Loading ...