36 UTD1000L User Manual
1. Press I button to set measurement type to DC Current, the measurement unit is μA, you
can use F3 to toggle betweenμA, mA and A ranges, the default is mA
2. Plug test lead connectors into input terminals(Black to COM and Red to μA/mA )
3. Then connect test leads to tested points, than take the DC current reading.
Figure 5-7 Measuring DC Current lower than 4Ma
To measure DC current lower than 400mA, do the following:
1. Press I button to set measurement type to DC Current,
2. Press F3 button to select mA range, andmeasurement unit will be mA .
3. Plug test lead connectors into input terminals(Black to COM and Red to μA/mA )
4. Then connect test leads to tested points, than take the DC current reading.
Figure 5-8 Measuring DC Current lower than 400mA
To measure DC current equal to or greater than 400mA, do the following:
1. Press I button to set measurement type to DC Current
2. Press F3 button to select A range, and measurement unit will be A .
3. Plug10A Current divider module (UT-M07) into input terminals COM and μA /mA, then
connect black and red test lead connectors respectively to the module (UT-M07)
4. Then connect test leads to tested points, than take the DC current reading.
Warning: If test leads are connected to COM and μA /mA without use of 10A Current divider
module (UT-M07) when measuring current greater than 400mA, the fuse in the product will be
burnt out. Please have the fuse replaced by qualified personnel if the incident happens.