30 UTD1000L User Manual
4.25 Autoset for Signals with DC Offset
UTD1000L Series is designed with powerful autoset function that makes it possible to
perform fast and accurate measurement in DCcoupled mode for signals with any DC
component.
Position parameter on upper right display of screen: This parameter stands for position
readout for the channel reference marker that moves from horizontal central line, if it is
positive value, it indicates the channel marker is above the horizontal central line and the
negative one is for the marker below the horizontal central line. To work out the final DC offset,
first move the waveform to the center of the screen, then multiply the Position value by
current amplitude range readout, that is final DC offset result you want.
For example:
Connect UTD1000CL channel with 70mVpp AC sine signal mixed with 1.1V DC offset, set the
channel to DC coupling, then perform AUTO function, the waveform in Figure 4-24A will show
on the screen. On the display, Position readout -53.63div indicates the channel reference
marker moves 53.63div downward from the horizontal central line. The current amplitude
range is 20mV/div, so DC offset readout: 20mV/div×53.63div =1.0726V
Vpp readout: 72mV means the AC peak-to-peak value of the signal
You can also directly read out the DC level through AVG value.
AVG value: 1.0726V, which stands for the signal DC level.
Through all parameters mentioned above, the signal can be understood very easily.
Note: For UTD1000DL models, CH1 and CH2 are both set to DC coupling when inputting the
signal above, then perform the AUTO function, after that, also move the waveform to the
center of the display(See Figure4-24B ), and then calculate DC offset with use of Position
readout and current amplitude range value.
Figure 4-23 UTD1000L Setup for Signals with DC Offset
As for traditional oscilloscope, it needs hard work to complete the task, for it has to set the
coupling to AC mode first to measure out the AC parameters, and then reset the coupling to