1-4
Specifications
Specifications apply at 18-28ºC after 30 minutes warm-up, at maximum output into 50
Waveforms
Standard Waveforms
Sine, square, triangle, DC, positive ramp, negative ramp, sin(x)/x, pulse, pulse train,
cosine, haversine and havercosine.
Sine, Cosine, Haversine, Havercosine
Range: 0·1 mHz to 16 MHz
Resolution: 0·1 mHz or 7 digits
Accuracy: 10 ppm for 1 year
Temperature stability: Typically <1 ppm/ºC.
Output level: 2.5 mV to 10 V p-p into 50
Harmonic distortion: <0.1% THD to 100 kHz
<–65 dBc to
20 kHz
<–50 dBc to 300 kHz
<–35 dBc to 10 MHz
<–30 dBc to 16 MHz
Non-harmonic spurious: <–65 dBc to 1 MHz,
<–65 dBc + 6
dB/octave 1 MHz to 16 MHz
Square
Range: 1 mHz to 16 MHz
Resolution: 1 mHz (4 digits)
Accuracy: ±1 digit of setting
Output level: 2.5 mV to 10 V p-p into 50
Rise and fall times: <25 ns
Triangle
Range: 0.1 mHz to 100 kHz
Resolution: 0.1 mHz or 7 digits
Accuracy: 10 ppm for 1 year
Output level: 2.5 mV to 10 V p-p into 50
Linearity error: <0.1 % to 30 kHz
Ramps and Sin(x)/x
Range: 0.1 mHz to 100 kHz
Resolution: 0.1 mHz (7 digits)
Accuracy: 10 ppm for 1 year
Output level: 2.5 mV to 10 V p-p into 50
Linearity error: <0.1 % to 30 kHz
Pulse and Pulse Train
Output level: 2.5 mV to 10 V p-p into 50
Rise and fall times: <25 ns
Period:
range:
resolution:
accura
cy:
100
ns to 100 s
4-digit
±1 digit of setting
1.888.610.7664 sales@GlobalTestSupply.com