The full procedure is as follows:
CAL 03 CH1 DC offset zero Adjust for 0 V ±5 mV
CAL 04 CH1 DC offset at + full scale Adjust for +10 V ±10 mV
CAL 05 CH1 DC offset at - full sca
le Check for -10 V ±3%
CAL 06 CH1 Multiplier zero Adjust for minimum voltage (AC)
CAL 07 CH1 Multiplier offset Adjust for 0 V ±5 mV
CAL 08 CH1 Waveform offset Adjust for 0 V ±5 mV
CAL 09 CH1 Output level at full-scale Adjust for 10 V ±10 mV
CAL 10 CH1 20 dB attenuator Adjust for 1 V ± 1 mV
CAL 11 CH1 40 dB attenuator Adjust for 0·1 V ±1 mV
CAL 12 CH1 10 dB attenuator Adjust for 2·236 V AC ±10 mV
CAL 13 CH1 Not used
CAL 14 CH1 Not used
CAL 15 CH1 Not used
CAL 16 CH2 DC offset zero Adjust for 0 V ±5 mV
CAL 17 CH2 DC offset at + full-scale. Adjust for +10 V ±10 mV
CAL 18 CH2 DC offset at - full-sca
le Check for -10 V ±3%
CAL 19 CH2 Multiplier zero Adjust for minimum voltage (AC)
CAL 20 CH2 Multiplier offset Adjust for 0 V ±5 mV
CAL 21 CH2 Waveform offset Adjust for 0 V ±5 mV
CAL 22 CH2 Output level at full-scale Adjust for 10 V ±10 mV
CAL 23 CH2 20 dB attenuator Adjust for 1 V ± 1 mV
CAL 24 CH2 40 dB attenuator Adjust for 0·1 V ±1 mV
CAL 25 CH2 10 dB attenuator Adjust for 2·236 V AC ±10 mV
CAL 26 CH2 Sum offset Adjust for 0 V ±5 mV
CAL 27 CH2 SCM level at full-scale Adjust for 5 V ±5 mV
CAL 28 CH2 AM level at full-scale Adjust for 10 V ±10 mV
CAL 29 CH3 DC offset zero Adjust for 0 V ±5 mV
CAL 30 CH3 DC offset at + full scale Adjust for +10 V ±10 mV
CAL 31 CH3 DC offset at - full sca
le Check for -10 V ±3%
CAL 32 CH3 Multiplier zero Adjust for minimum voltage (AC)
CAL 33 CH3 Multiplier offset Adjust for 0 V ±5 mV
CAL 34 CH3 Waveform offset Adjust for 0 V ±5 mV
CAL 35 CH3 Output level at full-scale Adjust for 10 V ±10 mV
CAL 36 CH3 20dB attenuator Adjust for 1 V ± 1 mV
CAL 37 CH3 40dB attenuator Adjust for 0·1 V ±1 mV
CAL 38 CH3 10dB attenuator Adjust for 2·236 V AC ±10 mV
CAL 39 CH3 Sum offset Adjust for 0 V ±5 mV
CAL 40 CH3 SCM level at full-scale Adjust for 5 V ±5 mV
CAL 41 CH3 AM level at full-scale Adjust for 10 V ±10 mV
CAL 42 CH4 DC offset zero Adjust for 0 V ±5 mV
CAL 43 CH4 DC offset at + full scale Adjust for +10 V ±10 mV
CAL 44 CH4 DC offset at - full sca
le Check for -10 V ±3%
CAL 45 CH4 Multiplier zero Adjust for minimum voltage (AC)
CAL 46 CH4 Multiplier offset Adjust for 0 V ±5 mV
CAL 47 CH4 Waveform offset Adjust for 0 V ±5 mV
CAL 48 CH4 Output level at full-scale Adjust for 10 V ±10 mV
CAL 49 CH4 20dB attenuator Adjust for 1 V ± 1 mV
CAL 50 CH4 40dB attenuator Adjust for 0·1 V ±1 mV
CAL 51 CH4 10dB attenuator Adjust for 2·236 V AC ±10 mV