Remote Operation
Remote Commands16
16-13
PULTRNWID <nrf1>,<nrf2> Set the width of pulse-train pulse number
<nrf1> to <nrf2> sec.
PULTRNDLY <nrf1>,<nrf2> Set the delay of pulse-train pulse number
<nrf1> to <nrf2> sec.
PULTRNMAKE Make the pulse-train and run it - similar to
the WAVE PULSTRN command.
ARB <cpd> Select an arbitrary waveform for output.
<cpd> must be the name of an existing
arbitrary
waveform. Backup memory is
always used as the source of the arb. The
arb will be copied to the channel memory if
necessary.
ARBLISTCH? Returns a list of all arbitrary waveforms in
the channel’s
memory. Each will return a
name and length in the following form
<cpd>,<nr1>. The list will end with
<rmt>.
ARBLIST? Returns a list of all arbitrary waveforms in
backup m
emory. Each will return a name
and length in the following form
<cpd>,<nr1>. The list will end with
<rmt>.
Arbitrary Waveform Create and Delete
Care should be take to ensure that all channels in the instrument are
running
in CONTINUOUS mode before using commands from this section.
Failure to observ
e this restriction may give unexpected results.
ARBDELETE <cpd> Delete the arbitrary waveform <cpd>
from backup memory.
ARBCLR <cpd> Delete the arb <cpd> from channel
mem
ory. The backup memory is not
changed.
ARBCREATE <cpd>,<nrf> Create a new, blank arbitrary waveform
with name <cpd> and length <nrf>
points.
ARBDEFCSV <cpd>,<nrf>,
<csv ascii data>
Define a new or existing arbitrary
wavefor
m with name <cpd> and length
<nrf> and load with the data in
<csv ascii data>.
If the arbitrary waveform does not exist it
will be creat
ed. If it does exist the length
will be checked against that specified and a
warning will be issued if they are different.
The edit limits will be set to the extremes
of the waveform
.
The data consists of ascii coded values, in
the range -2048 to +2047,
for each point.
The values are separated by a comma