Commissioning the dispensing pumps
Before the liquid dispensing pumps can be used the
liquid agent must be drawn up into the system and
the amount to be dispensed set.
Dispenser calibration
The dispensers need to be calibrated to ensure the
dispenser pumps supply the correct amount of liquid
agent.
To calibrate the dispensers with the machine
switched off:
^
Press and hold the rotary dial in.
^
Then switch the machine on with the "On" switch.
Dispenser calibration ~ Main menu
Dispenser pump 1
Dispenser pump 2 e
A
Selection choice: Dispenser pumps 1 to 13
After selecting a dispenser pump the following
message will appear in the display:
Dispenser calibration ~ Dispenser pump 1
V: 100ml t: 0s P: no
0/1 S A
The amount to dispense can be set between 0 and
9999 ml in 1 ml increments.
^
Press the 0/1 multi-function button and the
dispenser hose will start to fill. Press the 0/1 button
again to stop filling the hose.
^
Direct the hose of the dispenser pump being
calibrated into a suitable measuring flask.
,
Avoid contact with the skin and eyes, and
wear suitable protective clothing.
^
Press the A "Back" button to return to the main
menu.
Dispenser calibration ~ Main menu
Dispenser pump 1
Dispenser pump 2 e
A
^
Select dispenser pump.
Dispenser calibration ~ Dispenser pump 1
V: 100ml t: 0s P: no
0/1 S A
^
Press multi-function button 0/1.
To stop the pump press the 0/1 button again, or it
will switch off automatically after a period of 60
seconds.
^
Measure the amount collected in the measuring
flask.
Dispenser calibration ~ Dispenser pump 1
V: 100ml t: 30s P: 200.0 ml/min
0/1 S A
^ Enter this value in field "V:". Flow rate P in ml/min
will be calculated electronically.
^ Press the S "Save" button to save the values
measured in memory.
^ Reconnect the hose.
Cleaning and maintaining the dispensing system
The dispensing system should be flushed through
with clean warm water every two weeks and before
any longer down-times to prevent the risk of
blockages and corrosion.
^
The suction pipes should be cleaned through with
water.
^
Place the suction pipes in a container of hot water
(40 - 50°C).
^
Then use the dispenser calibration system to
operate each pump until the dispensing system is
thoroughly flushed through.
^
After reconnecting all parts check all connections,
hoses, dispensers and seals for leakages.