Binder FP 240 Drying and heating ovens

User Manual - Page 37

For FP 240.

PDF File Manual, 91 pages, Read Online | Download pdf file

Loading ...
Loading ...
Loading ...
FP (E1.1) 03/2019 page 37/90
Selection of controller menu language (Language)
German, English, or French can be selected.
Counter of operating hours (Oper.hs)
Information about the number of operating hours reached up to now or since the last reset. (no setting,
display only).
Max. number of operating hours (Op.limit)
Entry of a limit number of operating hours, i.e., the maximal number of operating hours that can be
run. Maximum setting: 9999. Reaching the limit has no effect.
Reset operating hours (Op.back)
Reset operating hours to zero.
Interface protocol (Protocol)
Modbus: The chamber interface can be used as a communication interface to connect it to a
computer. This serves to control the chamber by the APT-COM™ 4 Multi Management Software. It is
possible to read and write the values of all parameters.
Printer: A protocol printer for data printouts can be connected to the chamber interface. The printer
regularly protocols the actual temperature value with fixed formatting and with adjustable print
intervals.
In both cases an interface converter RS 422 / RS 232 is used.
Print interval (Prt.-Inv.)
Setting of the print interval in minutes. Function available only if setting Printerhas been selected in
the previous menu point.
Display illumination (Disp.LED)
Selection between continuous display illumination and limited illumination that will automatically go off
300 sec after the last entry.
Program type selection (PrgSelec)
Select between entry of two programs with up to 10 sections each or of one program with up to 20
sections.
When changing from 2 programs to 1 program or vice-versa, existing programs are
deleted in the program editor.
Maximum section duration (Prg.Time)
The maximum length of an individual program section can be set to either 99 hs 59 min or to 999 hs
59 min. This setting is then valid for all program sections.
When changing the maximum duration setting, pre-existing programs will be deleted in
the program editor.
Set-point programming type (Setp.sim)
Selection between “Ramp” and “Step”. With setting „Step“ selected, you don’t need to program the
transition section in the Program Editor.
If you select setting “step”, the controller will equilibrate only to constant temperatures;
programming ramps becomes impossible.
A change between settings “ramp” and “step” will influence all programs. This can
remarkably change the time courses of existing programs.