Selecting the next program sections to be entered
Display 1 shows
e.g. 02
(section S02 selected)
Display 2 shows
alternating
P01: --- SEC.
Section S02 has already been created.
enter new set-points for the individual parameters with X/W.
CONTINUE X/W
or:
Display 1 shows
e.g. 02
(section S02 selected)
Display 2 shows
alternating
P01: --- SEC.
Section S02 has not yet been created.
enter set-points for the individual parameters with X/W
NEW SEC. X/W
Select the next section to be entered with
arrow keys
Display 1 shows
e.g. 60.0
C
(actual temperature set-point)
Display 2 shows
alternating
S02:TEMP 60.0
(variable: temperature in °C)
CONTINUE X/W
(information: go on with X/W)
Enter the temperature set-point of S02 in °C
with arrow keys
Etc.
If all sections up to S10 or up to S20 have been programmed follows again section S01. In order to quit
the entry mode hit several times the EXIT button or wait 120 sec the controller returns to Normal
Display.
When changing the set-point, check the setting of the safety device (chap. 14.1).
If the fan is operated with less than 100 % speed, the temperature performance and the
spatial exactitude of the temperature can differ from the manufacturer’s specifications. The fan
speed rate should only be reduced if absolutely necessary to meet special requirements.