Chapter 2 Command System RIGOL
DSA800 Programming Guide 2-23
:CALCulate:MARKer<n>:STATe
Syntax
:CALCulate:MARKer<n>:STATe OFF|ON|0|1
:CALCulate:MARKer<n>:STATe?
Description
Enable or disable the specified marker.
Query the status of the specified marker.
Parameter
Name
Type
Range
Default
<n> Discrete 1|2|3|4 --
-- Bool OFF|ON|0|1 OFF|0
Explanation
If the specified marker is not enabled currently, it will be set to Normal marker (POSition) when this
command is sent to enable it.
Return Format
The query returns 0 or 1.
Example
The command below enables marker 1.
:CALCulate:MARKer1:STATe ON or :CALCulate:MARKer1:STATe 1
The query below returns 1.
:CALCulate:MARKer1:STATe?
:CALCulate:MARKer<n>:TRACe
Syntax
:CALCulate:MARKer<n>:TRACe <integer>
:CALCulate:MARKer<n>:TRACe?
Description
Set the trace to be marked by the specified marker.
Query the marker trace of the specified marker.
Parameter
Name
Type Range Default
<n> Discrete 1|2|3|4 --
<integer> Discrete 1|2|3|4 1
Explanation
<integer> denotes the trace selected and can be trace 1, trace 2, trace 3 or trace 4 (math trace). The trace
selected must have been enabled.
You can also use the
:CALCulate:MARKer<n>:TRACe:AUTO command to set the marker trace of the
specified marker to Auto.
Return Format
The query returns 1, 2, 3 or 4.
If the marker trace is set to Auto, the query returns the number of the trace marked by the marker.