DSA800 series programming guide - Page 200

For DSA815.

Loading ...
Loading ...
Loading ...
RIGOL Chapter 2 Command System
2-178 DSA800 Programming Guide
:SYSTem:LINemod:TYPe
Syntax
:SYSTem:LINemod:TYPe FACTory|USER1|USER2|USER3|USER4|USER5|USER6|OFF
Description
Set the preset setting used in line mode or exit line mode.
Parameter
Name
Type
Range
Default
-- Keyword
FACTory|USER1|USER2|USER3|USER4|
USER5|USER6|OFF
OFF
Example
The command below sets the preset setting used in line mode to USER4.
:SYSTem:LINemod:TYPe USER4
:SYSTem:LKEY
Syntax
:SYSTem:LKEY <license key>
:SYSTem:LKEY? <option>
Description
Install and activate the specified option.
Query the serial number of the specified option.
Parameter
Name
Type
Range
Default
<license key> ASCII STRING -- --
<option> Integer 1 to 4 --
Explanation
<option> denotes the number of the option. The numbers of the options and the corresponding order
numbers are as shown below.
Number of Option
1
2
3
4
Order Number of Option
AMK-DSA800
EMI-DSA800
VSWR-DSA800
PA-DSA875
Return Format
The query returns the serial number.
Example
The command below installs the option (the corresponding option number is 1) of which the serial number
is UADEYSACMA6RJACQTAWJZLLPBCVA.
:SYSTem:LKEY UADEYSACMA6RJACQTAWJZLLPBCVA
The query below returns UADEYSACMA6RJACQTAWJZLLPBCVA.
:SYSTem:LKEY? 0001
Loading ...
Loading ...
Loading ...