DSA800 series programming guide - Page 44

For DSA815.

Loading ...
Loading ...
Loading ...
RIGOL Chapter 2 Command System
2-22 DSA800 Programming Guide
:CALCulate:MARKer<n>[:SET]:STEP
Syntax
:CALCulate:MARKer<n>[:SET]:STEP
Description
Set the center frequency step of the spectrum analyzer to the frequency of the specified marker.
Parameter
Name
Type
Range
Default
<n> Discrete 1|2|3|4 --
Explanation
If the specified marker is Normal marker, the center frequency step will be set to the frequency of the
marker.
If the specified marker is Delta, Delta Pair or Span Pair marker, the center frequency step will be set to the
frequency of the delta marker.
This command is only available when the specified marker is enabled.
This function is invalid in zero span mode.
Example
The command below sets the center frequency step of the spectrum analyzer to the frequency of marker 4
(Normal marker).
:CALCulate:MARKer4:SET:STEP
:CALCulate:MARKer<n>[:SET]:STOP
Syntax
:CALCulate:MARKer<n>[:SET]:STOP
Description
Set the stop frequency of the spectrum analyzer to the frequency of the specified marker.
Parameter
Name
Type
Range
Default
<n> Discrete 1|2|3|4 --
Explanation
If the specified marker is Normal marker, the stop frequency will be set to the frequency of the marker.
If the specified marker is Delta, Delta Pair or Span Pair marker, the stop frequency will be set to the
frequency of the delta marker.
This command is only available when the specified marker is enabled.
This function is invalid in zero span mode.
Example
The command below sets the stop frequency of the spectrum analyzer to the frequency of marker 2
(Normal marker).
:CALCulate:MARKer2:SET:STOP
Loading ...
Loading ...
Loading ...