DSA800 series programming guide - Page 162

For DSA815.

Loading ...
Loading ...
Loading ...
RIGOL Chapter 2 Command System
2-140 DSA800 Programming Guide
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]
Syntax
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer] <ampl>
[:SENSe]:POWer[:RF]:MIXer:RANGe[:UPPer]?
Description
Set the maximum power of the input mixer.
Query the maximum power of the input mixer.
Parameter
Name
Type
Range
Default
<ampl> Integer -30 dBm to 0 dBm -10 dBm
Return Format
The query returns the maximum power of the input mixer in scientific notation and the unit is dBm.
Example
The command below sets the maximum power of the input mixer to -20 dBm.
:SENSe:POWer:RF:MIXer:RANGe:UPPer -20
The query below returns -2.000000E+01.
:SENSe:POWer:RF:MIXer:RANGe:UPPer?
[:SENSe]:SIGCapture[:STATe]
Syntax
[:SENSe]:SIGCapture[:STATe] OFF|ON|0|1
[:SENSe]:SIGCapture[:STATe]?
Description
Turn the Sig Capture function on or off.
Query the on/off status of the Sig Capture function.
Parameter
Name
Type
Range
Default
——
Bool OFF|ON|0|1 OFF|0
Return Format
The query returns 0 or 1.
Example
The command below turns the Sig Capture function on.
:SENSe:SIGCapture:STATe ON or :SENSe:SIGCapture:STATe 1
The query below returns 1.
:SENSe:SIGCapture:STATe?