DSA800 series programming guide - Page 168

For DSA815.

Loading ...
Loading ...
Loading ...
RIGOL Chapter 2 Command System
2-146 DSA800 Programming Guide
[:SENSe]:SIGCapture:2FSK:MARK1:Switch[:STATe]
Syntax
[:SENSe]:SIGCapture:2FSK:MARK1:Switch[:STATe] OFF|ON|0|1
[:SENSe]:SIGCapture:2FSK:MARK1:Switch[:STATe]?
Description
Turn the Marker 1 on or off.
Query the on/off of Marker 1.
Parameter
Name
Type
Range
Default
——
Bool OFF|ON|0|1 OFF|0
Explanation
This command is valid only when 2FSK is turned on.
Return Format
The query returns 0 or 1.
Example
The command below turns the Marker 1 on.
:SENSe:SIGCapture:2FSK:MARK1:Switch:STATe ON or :SENSe:SIGCapture:2FSK:MARK1:Switch:STATe 1
The query below returns 1.
:SENSe:SIGCapture:2FSK:MARK1:Switch:STATe?
[:SENSe]:SIGCapture:2FSK:MARK2[:FREQ]
Syntax
[:SENSe]:SIGCapture:2FSK:MARK2[:FREQ] <freq>
[:SENSe]:SIGCapture:2FSK:MARK2[:FREQ]?
Description
Set the frequency of Marker 2.
Query the frequency of Marker 2.
Parameter
Name
Type
Range
Default
<freq>
Consecutive Real
Number
(Center Freq -Span/2) to (Center Freq + Span/2)
Wherein, Span fixes as 1.5 MHz
Center Freq +
Span/2
Explanation
This command is valid only when 2FSK is turned on.
Return Format
The query returns the frequency of Marker 2 in integer and the unit is Hz.
Example
The command below sets the frequency of Marker 2 to 750 MHz.
:SENSe:SIGCapture:2FSK:MARK2:FREQ 750000000
The query below returns 750000000.
:SENSe:SIGCapture:2FSK:MARK2:FREQ?
Loading ...
Loading ...
Loading ...