DSA800 series programming guide - Page 147

For DSA815.

Loading ...
Loading ...
Loading ...
Chapter 2 Command System RIGOL
DSA800 Programming Guide 2-125
[:SENSe]:EBWidth:MAXHold:STATe
Syntax
[:SENSe]:EBWidth:MAXHold:STATe OFF|ON|0|1
[:SENSe]:EBWidth:MAXHold:STATe?
Description
Enable or disable the Max Hold.
Query the status of the Max Hold.
Parameter
Name
Type
Range
Default
-- Bool OFF|ON|0|1 OFF|0
Explanation
This command is only available when the emission bandwidth measurement is enabled.
When Max Hold is enabled, each measurement result is compared with the previous result and the
maximum is displayed.
When Max Hold is disabled, the current measurement result is displayed.
Max Hold and the average measurement function are mutually exclusive and the average measurement
function will be automatically disabled when Max Hold is enabled.
Return Format
The query returns 0 or 1.
Example
The command below enables the Max Hold.
:SENSe:EBWidth:MAXHold:STATe ON or :SENSe:EBWidth:MAXHold:STATe 1
The query below returns 1.
:SENSe:EBWidth:MAXHold:STATe?
[:SENSe]:EBWidth:XDB
Syntax
[:SENSe]:EBWidth:XDB <real_amp>
[:SENSe]:EBWidth:XDB?
Description
Set the value of X dB of the EBW measurement.
Query the value of X dB of the EBW measurement.
Parameter
Name
Type
Range
Default
<real_amp>
Consecutive Real Number
-100 dB to -0.1 dB
-10 dB
Explanation
This command is only available when the emission bandwidth measurement is enabled.
Return Format
The query retunes the value of X dB in scientific notation.
Loading ...
Loading ...