RIGOL Chapter 2 Command System
2-156 DSA800 Programming Guide
[:SENSe]:VSWR:NREFlect
Syntax
[:SENSe]:VSWR: NREFlect
Description
The spectrum analyzer executes the second measurement of the VSWR measurement.
Explanation
This command is only available when the VSWR measurement is enabled.
Disconnect the device under test and send the [:SENSe]:VSWR:FREFlect command; then connect the the
device under test and send this command.
[:SENSe]:VSWR:RESet
Syntax
[:SENSe]:VSWR:RESet
Description
Restore the parameters of the VSWR measurement to their initial states.
[:SENSe]:VSWR:STATe
Syntax
[:SENSe]:VSWR:STATe OFF|ON|0|1
[:SENSe]:VSWR:STATe?
Description
Enable or disable the VSWR measurement.
Query the status of the VSWR measurement.
Parameter
Name
Type Range Default
-- Bool OFF|ON|0|1 OFF|0
Return Format
The query returns 0 or 1.
Example
The command below enables the VSWR measurement.
:SENSe:VSWR:STATe ON or :SENSe:VSWR:STATe 1
The query below returns 1.
:SENSe:VSWR:STATe?