2-108 DSA800 Programming Guide
[:SENSe]:BANDwidth:VIDeo:RATio
Syntax
[:SENSe]:BANDwidth:VIDeo:RATio <number>
[:SENSe]:BANDwidth:VIDeo:RATio?
Description
Set the V/R ratio, namely the ratio of VBW to RBW.
Query the V/R ratio.
Parameter
Name
Type Range Default
<number> Discrete 0.000001 to 30000, at 1-3-10 step 1
Return Format
The query returns the V/R ratio in scientific notation.
Example
The command below sets the V/R ratio to 0.01.
:SENSe:BANDwidth:VIDeo:RATio 0.01
The query below returns 1.000000E-02.
:SENSe:BANDwidth:VIDeo:RATio?
[:SENSe]:CHPower:AVERage:COUNt
Syntax
[:SENSe]:CHPower:AVERage:COUNt <integer>
[:SENSe]:CHPower:AVERage:COUNt?
Description
Set the number of averages of the channel power measurement.
Query the number of averages of the channel power measurement.
Parameter
Name
Type
Range
Default
<integer> Integer 1 to 1000 10
Explanation
This command is only available when the channel power measurement is enabled.
Return Format
The query returns the number of averages in integer.
Example
The command below sets the number of averages to 100.
:SENSe:CHPower:AVERage:COUNt 100
The query below returns 100.