DSA800 series programming guide - Page 35

For DSA815.

Loading ...
Loading ...
Loading ...
Chapter 2 Command System RIGOL
DSA800 Programming Guide 2-13
:CALCulate:MARKer:FCOunt[:STATe]
Syntax
:CALCulate:MARKer:FCOunt[:STATe] OFF|ON|0|1
:CALCulate:MARKer:FCOunt[:STATe]?
Description
Enable or disable the frequency counter.
Query the status of the frequency counter.
Parameter
Name
Type
Range
Default
-- Bool OFF|ON|0|1 OFF|0
Explanation
Use the
:CALCulate:MARKer:FCOunt:X? command to query the reading of the frequency counter.
Return Format
The query returns 0 or 1.
Example
The command below enables the frequency counter.
:CALCulate:MARKer:FCOunt:STATe ON or :CALCulate:MARKer:FCOunt:STATe 1
The query below returns 1.
:CALCulate:MARKer:FCOunt:STATe?
:CALCulate:MARKer<n>:CPEak[:STATe]
Syntax
:CALCulate:MARKer<n>:CPEak[:STATe] OFF|ON|0|1
:CALCulate:MARKer<n>:CPEak[:STATe]?
Description
Enable continuous peak search and mark the peak using the specified marker or disable continuous peak
search.
Query the status of continuous peak search.
Parameter
Name
Type
Range
Default
<n>
Discrete
1|2|3|4
1
--
Bool
OFF|ON|0|1
OFF|0
Return Format
The query returns 0 or 1.
Example
The command below enables continuous peak search and marks the peak using marker 2.
:CALCulate:MARKer2:CPEak:STATe ON or :CALCulate:MARKer2:CPEak:STATe 1
The query below returns 1.
:CALCulate:MARKer2:CPEak:STATe?