DSA800 series programming guide - Page 155

For DSA815.

Loading ...
Loading ...
Loading ...
Chapter 2 Command System RIGOL
DSA800 Programming Guide 2-133
Example
The command below sets the average mode to repeat average.
:SENSe:HDISt:AVERage:TCONtrol REPeat
The query below returns REP.
:SENSe:HDISt:AVERage:TCONtrol?
[:SENSe]:HDISt:NUMBers
Syntax
[:SENSe]:HDISt:NUMBers <integer>
[:SENSe]:HDISt:NUMBers?
Description
Set the number of the harmonics to be measured.
Query the number of the harmonics to be measured.
Parameter
Name
Type
Range
Default
<integer>
Integer
2 to 10
10
Explanation
This command is only available when the harmonic distortion measurement is enabled.
Return Format
The query returns the number of harmonics in integer.
Example
The command below sets the number of harmonics to 5.
:SENSe:HDISt:NUMBers 5
The query below returns 5.
:SENSe:HDISt:NUMBers?
[:SENSe]:HDISt:TIME
Syntax
[:SENSe]:HDISt:TIME <time>
[:SENSe]:HDISt:TIME?
Description
Set the sweep time of the harmonic distortion measurement.
Query the sweep time of the harmonic distortion measurement.
Parameter
Name
Type Range Default
<time>
Consecutive Real Number 20 us to 7.5 ks 37.5 ms
Explanation
This command is only available when the harmonic distortion measurement is enabled.
Return Format
The query returns the sweep time in scientific notation and the unit is s.
Loading ...
Loading ...