DSA800 series programming guide - Page 110

For DSA815.

Loading ...
Loading ...
Loading ...
RIGOL Chapter 2 Command System
2-88 DSA800 Programming Guide
:OUTPut Subsystem
Command List:
:OUTPut[:STATe]
:OUTPut[:STATe]
Syntax
:OUTPut[:STATe] OFF|ON|0|1
:OUTPut[:STATe]?
Description
Enable or disable the output of the tracking generator.
Query the on/off status of the tracking generator.
Parameter
Name
Type
Range
Default
--
Bool
OFF|ON|0|1
OFF|0
Explanation
This command is only available for DSA815-TG/DSA832-TG/DSA875-TG.
Return Format
The query returns 0 or 1.
Example
The command below enables the output of the tracking generator.
:OUTPut:STATe ON or :OUTPut:STATe 1
The query below returns 1.
:OUTPut:STATe?