DSA800 series programming guide - Page 145

For DSA815.

Loading ...
Loading ...
Loading ...
Chapter 2 Command System RIGOL
DSA800 Programming Guide 2-123
[:SENSe]:EBWidth:AVERage[:STATe]
Syntax
[:SENSe]:EBWidth:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:EBWidth:AVERage[:STATe]?
Description
Enable or disable the average measurement function of the emission bandwidth measurement.
Query the status of the average measurement function of the emission bandwidth measurement.
Parameter
Name
Type
Range
Default
-- Bool OFF|ON|0|1 OFF|0
Explanation
This command is only available when the emission bandwidth measurement is enabled.
Return Format
The query returns 0 or 1.
Example
The command below enables the average measurement function.
:SENSe:EBWidth:AVERage:STATe ON or :SENSe:EBWidth:AVERage:STATe 1
The query below returns 1.
:SENSe:EBWidth:AVERage:STATe?
[:SENSe]:EBWidth:AVERage:TCONtrol
Syntax
[:SENSe]:EBWidth:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:EBWidth:AVERage:TCONtrol?
Description
Set the average mode of the emission bandwidth measurement.
Query the average mode of the emission bandwidth measurement.
Parameter
Name
Type
Range
Default
-- Keyword EXPonential|REPeat EXPonential
Explanation
EXPonential: exponential average
REPeat: repeat average
When exponential average is selected, the result is the exponential average of the latest N (specified by the
[:SENSe]:EBWidth:AVERage:COUNt command) measurement results.
When repeat average is selected, the result is the arithmetic average of the latest N (specified by the
[:SENSe]:EBWidth:AVERage:COUNt command) measurement results.
This command is only available when the emission bandwidth measurement is enabled.
Return Format
The query returns EXP or REP.