Troubleshooting
14.2 Comments 14
14-3
AC voltage present on the chassis Use a wall plug tester to check the main
power plug at the wall. Use an ohmmeter to
check the continuity between the ground
prong on the PEM and the chassis. If the
resistance reading is greater than three
ohms, there is a problem. Check the power
cord for continuity on the ground prongs. If
the resistance is greater than one ohm,
replace the power cord.
14.2 Comments
14.2.1 EMC Directive
Hart Scientifics’ equipment has been tested to meet the European
Electromagnetic Compatibility Directive (EMC Directive, 89/336/EEC). The
Declaration of Conformity for your instrument lists the specific standards to which
the unit was tested.
14.2.2 Low Voltage Directive (Safety)
In order to comply with the European Low Voltage Directive (73/23/EEC), Hart
Scientific equipment has been designed to meet the EN 61010-1 and EN 61010-
2-010 standards.