FLUKE 7102-156 Liquid Bath Calibrator, -5 to 125°C

User Manual for the Fluke 7102-156 Liquid Bath Calibrator, -5 to 125°C - Page 48

For 7102-156.

PDF File Manual, 68 pages, Read Online | Download pdf file

7102-156 photo
Loading ...
Loading ...
Loading ...
7102
User's Guide
9-8
The proportional band width is easily adjusted from the front panel. The width
may be set to discrete values in degrees C or F depending on the selected units.
The proportional band adjustment is be accessed within the secondary menu.
Press “SET” and “EXIT” to enter the secondary menu and show the heater
power. Then press “SET” to access the proportional band.
+ Access heater power in secondary menu
12.0 P Heater power in percent
Access proportional band
ProP Flashes “ProP” and the setting
5.0 Proportional band setting
To change the proportional band press “UP” or “DOWN”.
4.0 New proportional band setting
To store the new setting press “SET”. Press “EXIT” to continue without storing
the new value.
Accept the new proportional band setting
9.8 Controller Configuration
The controller has a number of configuration and operating options and
calibration parameters which are programmable via the front panel. These are
accessed from the secondary menu after the proportional band function by
pressing “SET”. Pressing “SET” again enters the first of three sets of
configuration parameters: operating parameters, serial interface parameters, and
calibration parameters. The menus are selected using the “UP” and “DOWN”
keys and then pressing “SET”. (See Figure 7 on page 32.)
9.8.1 Operating Parameters
The operating parameters menu is indicated by,
PAr Operating parameters menu
The operating parameters menu contains the High Limit and Stir Speed
parameters.