10.2 Interface Commands
The various commands for accessing the calibrator functions via the digital
interfaces are listed in this section (see Table 5). These commands are used with
the RS-232 serial interface. The commands are terminated with a carriage-return
character. The interface makes no distinction between upper and lower case
letters, hence either may be used. Commands may be abbreviated to the
minimum number of letters which determines a unique command. A command
may be used to either set a parameter or display a parameter depending on
whether or not a value is sent with the command following a “ =” character. For
example “s” <CR > will return the current set-point and “s =150.0”<CR> will set
the set-point to 150.0 degrees.
In the following list of commands, characters or data within brackets, “[” and “]”,
are optional for the command. A slash, “/”, denotes alternate characters or data.
Numeric data, denoted by “n”, may be entered in decimal or exponential notation.
Characters are shown in lower case although upper case may be used. Spaces
may be added within command strings and will simply be ignored. Backspace
(BS, ASCII 8) may be used to erase the previous character. A terminating CR is
implied with all commands.
Table 5. Controller Communications Commands